| US$200.00-800.00 | 1 Piece (MOQ) |
| Model | AD-12 |
| Application Range | Adult/Pediatric/Neonatal/Medicin e/Surgery/ Operating Room/ICU/CCU |
| Display | 12.1* real color TFT screen |
| working environment conditions Temperature | Working temperature0-40ºC Working humidity≤85% |
| Power supply voltage | AC 100~240V AC |
| power frequency | 50/60Hz, maximum input power 70VA |
| Built-in lithium battery | d.c11.1V not |
| Product Size | 262*336*84mm |
| product weight | 2.72kg |
| Gain | X0.25,x0.5, x1,X2,Auto |
| Waveform speed | 12.5mm/s,25mm/s,50mm/s |
| Diagnosticmode | 0.05-130Hz Monitoring mode:0.5-40Hz Operation mode:1~20Hz |
| Common Mode Rejection Ratio | Diagnosticmode:>90dB Monitoring mode:>100dB Operation mode:>100dB |
| Electrode polarization voltage | ±300mV Input impedan ce≥5M |
| DC biasvoltage | ±300mV |
| Calibration signal | 1mV (peak-to-pea k), accuracy±5% |
| ST segment detection measurement: Measurement range | -2.0mV-+2.0mV |
| Measuring Range and Precision Range | Adult Systolic blood pressure40~270mmHg Diastolic blood pressure 10~215mmHg Mean blood pressure 20~235mmHg Static pressure range 0~300mmHg Static pressure precision ±3mmHg |
| Pressure precision | Max. average error: ±5mmHgi Max stand ard devaton: 8mmHg |
| Spo2 | Measuringrange0~100% Resolution 1% The accuncyis within the range of 70% to 100%, and the measurement error should be ±2%; In therange of 0% to 69%, the measurement error is undefined |
| Pulse Rate | Measuringrange 30~300bpm Resolution 1bpm Accuracy±3bpm Updatetime ls |
| Standard accessories | NIBP cuff&tube ECG cable & electrodes Sp02 sensor TEMP probe Lithium-ion battery Power cable Operator's manual |
| Optional accessories | End-tidal CO2 (optional) Recorder (optional) Dual IBP Nellcor (optional) |
| Store & Reviewing | |
| ECG | 30 minutes one important lead's ECG waveform |
| Alarm | 1800groups Alarm events reviewing |
| NIBP | 400 groups NIBP measurement |
| Arrhythmia | 128 groups data (8 seconds ECG waveforms) |
| Power | off storage:72 hours trend data & 1 ECG wave form(option) |
| ECG | |
| Mode | 5-leads (standard);3-leadsor 12-leads (option) |
| Lead selection | I, 11, 1H,aVR,aVL,aVF,V1~V6(option) |
| Gain | AUTO,0.25x,0.5x,1.0x,2.0x,4.0x -Insulation Breakdown Voltage4000VAC 50/60Hz Sweep speed 6.25mm/s,12.5mm/s,25mm/s,50mm/s |
| HR Range | Adult:10~300bpm Child & Neonate:10~350bpm -HRAccuracy ±1%or±1 bpm,whichever is greater |
| IBP(option) | |
| Channel | 2 |
| Measurement Range | -50~+300mmH |
| Unit | mmHg,kPa |
| Accuracy | ±2mmHgor2%,whichever is greater |
| Alarm | |
| Three Level | Low, medium and high |
| Indication | Auditory and visual |
| Setup | Default and custom |
| Silence | All alarms can be silenced |
| Volume | 45-85 dB measured at 1 meter |
| Medical Equipment Portable Capnograph SpO2 Saturation Vital Signs Monitor Handheld Vital Signs Monitoring Equipment | Medical Equipment Portable Plug-in Patient Monitor with 15′ ′ Color TFT Screen | Medical Equipment 3 Channel 12 Lead Abnormal Data Alarm ECG EKG Machine with Analyzer |
| Hospital Syringe Pump Syringe Professional Portable Best Selling Syringe Pump | Medical Automatic High Precision Peristaltic IV Volumetric Portable Infusion Pump | Medical Equipment 10.2 Inch Touch Screen Portable Surgical ICU Anesthesia Machine with 2 Vaporizers for Operation Room |
| Medical Equipment ICU Invasive Surgical Ventilator with 10.4" TFT Touch Screen and Air Compressor | Accurate Medical Oximeters Adult Fingertip Neonatal Oxymetre Display Finger Pulse Oximeters Digital | Color Doppler Handheld Wireless Ultrasound Scanner USB&Wi-Fi 3 in 1 Probe Ultrasound Device Medical Diagnostic |